Hegar Straight Needle Holder - 150 mm

Hegar Needle Holder Straight, length 150 mm (6"), heavy pattern design with serrated jaws and longitudinal groove, suitable for sutures up to 3/0. Reusable surgical instrument, non-sterile and latex-free. Ideal for procedures on thicker or tougher tissues, ensures secure grip of the thread and optimal control. Sterilizable by fractional steam at 134°C.

€93.00
VAT excluded
Quantity

 

Security policy (edit with Customer reassurance module)

 

Delivery policy (edit with Customer reassurance module)

 

Return policy (edit with Customer reassurance module)

HEGAR STRAIGHT NEEDLE HOLDER HEAVY PATTERN, SERRATED WITH GROOVE - 150 MM

The Hegar 150 mm Straight Needle Holder – Heavy Pattern is designed for suturing thicker or tougher tissues, with threads up to 3/0, providing maximum stability and control during surgical procedures.

Technical features:

  • Geometry: straight
  • Total length: 150 mm (6")
  • Serrated jaws with longitudinal groove
  • Heavy pattern design for greater robustness
  • Ring handle
  • Indicated for sutures up to 3/0
  • Reusable instrument
  • Direct UDI marking
  • Latex-free
  • Standard version: BM065R
  • Non-sterile

The heavy pattern allows for a firm grip even on thicker threads and tough tissues, making the Hegar needle holder suitable for general, orthopedic, and surgical procedures on adult patients.

Areas of application:

General surgery, orthopedics and traumatology, plastic surgery, vascular surgery, cardio-thoracic surgery, neurosurgery, maxillofacial surgery, and dentistry.

Reconditioning and sterilization

Manual cleaning:

  • Immersion with chemical disinfection for 15 minutes (e.g., B. Braun Stabimed 2% solution) with brushing.

Mechanical cleaning:

  • Alkaline washing with thermal disinfection at 90°C for 5 minutes according to the thermal disinfectant program.

Validated sterilization:

  • Fractionated steam under vacuum at 134°C with a holding time of 5 minutes.

Note:

  • Instruments with joints or friction surfaces must be lubricated before each sterilization cycle to prevent corrosion.

Product Details

BM245R