Aesculap-bbraun

STEVENS Dissection Scissors (Tenotomy) Curved – 110 mm, blunt tips

STEVENS Tenotomy type dissection scissors, curved, 110 mm, with blunt/blunt tips and delicate design. Reusable, latex-free, with direct UDI marking and supplied non-sterile. Ideal for dental, ENT, vascular, urological, plastic, pediatric, orthopedic, neurosurgery, maxillofacial, hand-foot, cardiothoracic, and general surgery.

€112.00
VAT excluded
Quantity

 

Security policy (edit with Customer reassurance module)

 

Delivery policy (edit with Customer reassurance module)

 

Return policy (edit with Customer reassurance module)

STEVENS CURVED DISSECTION SCISSORS - 110 mm

The STEVENS (Tenotomy) curved dissection scissors are precision surgical instruments designed for atraumatic dissections and controlled maneuvers on delicate tissues.

The curvature of the blades facilitates access to deep anatomical planes and improves visibility of the surgical field, while the blunt/blunt tips reduce the risk of accidental injuries. The delicate design and compact length ensure maximum maneuverability and control.

Technical features:

  • Model: STEVENS
  • Type: Tenotomy
  • Geometry: curved
  • Tip profile: blunt/blunt
  • Design: delicate
  • Length: 110 mm (4 1/4")
  • Handle: ring-shaped
  • Quality standard: Classic
  • Direct UDI marking: present
  • Reusable
  • Non-sterile
  • Latex-free
  • Without tungsten carbide inserts

Areas of application:

Dental surgery, ENT, vascular, urological, plastic, pediatric, orthopedic/traumatology, neurosurgery, maxillofacial, hand-foot surgery, cardiothoracic, and general/visceral surgery.

Cleaning and disinfection

Manual treatment:

  • Disinfection by immersion and cleaning with a brush
  • Chemical disinfection: 15 minutes with Stabimed 2% solution
  • Drying: not applicable

Mechanical treatment:

  • Mechanical alkaline cleaning with thermal disinfection
  • Thermal disinfection: 90°C for 5 minutes
  • Drying according to the program of the thermal disinfectant (Digit 4 RDS Material)

Sterilization

Validated steam process:

  • Fractionated vacuum at 134°C
  • Holding time: 5 minutes

Technical note:

To prevent corrosion from friction, articulated instruments or those with metal surfaces subject to rubbing must be lubricated before each sterilization cycle.

Product Details

BC173R